1. Application
Monitoring of pH in many different industrial processes to reduce cost and optimize yield, especially in harsh conditions such as municipal wastewater and industrial wastewater.
• Water/wastewater treatment
•Inflow of the plant, water collection well, aeration tank, the effluent, etc.
• Pulp and paper, power plant, mine
•Water treatment, chemical treatment feeding
2. Features and Benefits
• The maintenance of electrode is easy and low cost, the reference system repels dirt and rarely clog.
• No electrolyte refilling is required, it can be used under the condition of the max pressure 6bar.
• Strong anti-interference by sulfur.
• History curve function: the measurement data is collected every 5 minutes, data can be stored for one month.
• Notepad function, provides the record of operating and alarming time.
• When the electrode is calibrated, the mode, time and results are recorded, easy to review and analysis the change.
3. Specifications
(1) Performance characteristics
Measuring Range: 0~*4pH, **0~**0°C
Resolution: 0.*1pH, 0.1°C
Accuracy: ±0.*5pH
Repeatability: 1%
Sensitivity: 0.*2pH
Temperature compensation: **0~**0°C,automatic
Calibration: Factory calibrated, also can be calibrated by customer on site
Display: LCD displays data including date, time, measurement value, trendline, etc.
Data Storage: Automatic data store, power-off data protection
(2) Outputs
Analog Outputs: ***0mA/***0 mA output( maximum load **0 ohms)
Relay Outputs: 2 Relays with programmable hysteresis
Relay Capacity: 3A, **0VAC
Digital Communication: MODBUS RS**5,Profibus DP is optional
(3) Power Supply
AC Power Supply: *0~**5 VAC, *0Hz
DC Power Supply: *4 VDC ±*0%
(4) Physical parameters
Dimension of Transmitter: **6×**6×**8mm(W×H×D) (5.7*5.7*4.3 in.)
Dimension of Sensor: Φ*8×**5mm (1.1*6.5 in.)
Materials: *-Transmitter: PC
*-Sensor: Polyester
Process connection of sensor: NPT 3/4
Weight: *-Transmitter: 1Kg
*-Sensor: 0.5Kg
Length of sensor’s Cable: *0m, Shielded cable (other length available)
(5) Environmental conditions
Ingress protection: *-IP*6 (transmitter)
*-IP*8 (Sensor)
Operating Temperature: *-Transmitter: **0~*0°C
*-Sensor: 0~*0°C
Operating Pressure: ≤6Bar
4. Ordering Information
(1) Transmitter Selection
TPH A B CD
*- A, single or duplex isolated 4 ~ *0mA output:
Code Meaning
1 Single isolated 4 ~ *0mA output (pH value)
2 Duplex isolated 4 ~ *0mA output (pH and temperature measured value)
*- B, Signal Output:
Code Meaning
0 4~*0mA
1 RS **5 (Modbus)
2 Profibus DP
*- CD, PowerVersion:
Code Meaning
AC **0VAC
DC *4 VDC
Example
TPH*0AC, pH analyzer transmitter, output Single isolated 4 ~ *0mA (pH), **0V AC Power Supply
(2) Sensor Selection
TPH – S A CXX
*- A, Usage:
Code Meaning
0 Adapted to sewage
1 Other purposes
*- CXX, Cable Length:
Code Meaning
C*0 *0m cable
CXX XX m cable
Example
TPH-S0C*0, a pH sensor,adapted to sewage, with *0m cable.
(3) Accessories
FA***0: Immersion mounting bracket, stainless steel