La description
â– Main
body                Â
High-strength engineering plastic rack, light,
beautiful and corrosion resistant
â– Scope of
application         Â
***5kg Animal
â– Gas
source                Â
O2: 0.*7~0.*5MPa   N2O: 0.*7~0.*5MPa
â– Flow
meter         Â
       O2:
0.1~1.0L/min  1.1~*0L/min
(N*0: 0.1~1.0L/min  1.1~*0L/min)
â– O2, N2O linkage and N2O Stopper
  When using nitrous oxide, oxygen
concentration>*5%;
When the oxygen pressure<0.2Kpa, the flow of nitrous oxide
would be cut off.
â– Flow rate of rapid oxygen supply *5~*5L/min
â– Low oxygen pressure alarm    There
will be sound alarm when the oxygen
pressure < 0.2MPa
â– Vaporizer            Â
     (it has the function of automatic
compensation based on pressure, temperature, and flow rate.The
regulation range of evaporator concentration is 0~5 vol%.)
Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can
be chosen for application as required by the customer.
â– Respiratory
circuit       Â
    working mode: all-close, semi-close,
semi-open
       Â
                    releasing
pressure: 1~4 kPa
â– Respiratory bellows  Â
        bellows for
adults, bellows for children
tidal volume range:0~***0 ml
Â
Part ofVentilator
â– Display
mode             Â
  High-definition 4.8 LCD screen display
â– VentilationÂ
mode       Â
    IPPV, MANUAL (Manual monitoring of tidal
volume, ventilation volume, respiratory rate)
â– Ventilation
parameters       Â
 Tidal
volume       Â
*0~***0ml
Rate             Â
 4~*0 bpm
I:EÂ Â Â Â Â Â Â Â Â Â Â Â Â
  2:1~1:4
Pressure range  Â
    1.0~6.0 kPa
â– Parameters for ventilation monitoring  tidal
volume, ventilation volume, IPPV rate, I/E, peak pressure of
airway, pressure - time waveform
â– Security Alarm System
Airway pressure alarmÂ
        upper limit
setting range 1.0~6.0 kPa
low limit setting range  0.4~2.0 kPa
TV
alarm                          Â
upper limit setting range 0.5~2.0L/min
low limit setting range  0.*2~1.0L/min
Sustained high-pressure alarm     it
will give alarm when stress have consistently
been higher than 2.5 kPa
Suffocation
alarm           Â
 it will give sound and light alarm if there is
no tidal volume  in *5s
Power alarm
â– Power           Â
          Â
  AC **0V±*0%     *0Hz±1Hz
â– (Specific interface of waste anesthesia gas)
Â
*Items within brackets are optional.